FM flowmeters

The FM flowmeters are devices for the supply of medical gases having from two to six antistatic and graduated tubes and completed with an integrated dosage unit. 

STRUCTURE
The body is made of anodized aluminum and the inlet connections for O2, Air and N2O at the bottom of the flowmeter are arranged to be connected to hospital centralized gas plants or to cylinders complete with pressure regulators.

WORKING PRINCIPLE AND SAFETY
The gas feeding unit includes a safety valve automatically blocking the N2O supply when the feeding pressure of O2 drops below the value of approximately 0.4 bar. The mixtures of O2 and N2O can be supplied to the patient through a dosage unit. An incorporated mixing device proportions the N2O in such a way that a minimum of at least 30% of O2 is always present in the gas mixture. At the outlet of the dosage unit, an anesthesia evaporator can be connected or the gas mixture can be supplied directly to the patient through the gas feeding unit. The O2+ flush button can be used for a quick supply of O2 directly at the outlet of the mixture on the gas feeding unit.

read more

Tech Info

 

FM 2200

FM 2300

FM 2500

Sizes (LxWxH)

148x143x318 mm

148x143x318 mm

148x143x318 mm

Weight

5.52 Kg

5.57 Kg

5.64 Kg

Dosage range

O2: 0.2 - 15 L/min.
N2O: 0.2 - 12L/min.

O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.

O2: 0.2 - 15 L/min.
Air: 0.2 - 15 L/min.
N2O: 0.2 - 12 L/min.

 

FM 2800

FM 2900

Sizes (LxWxH)

148x143x318 mm

148x190x318 mm

Weight

5.69 Kg

7.2 Kg

Dosage range

O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
Air:  0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.

O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
Air:  0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.
CO2 0.1 - 2 L/min. (limited at 1 L/min.)

TECHNICAL SPECIFICATIONS

Accuracy

±10% read value between 10% (or ±0.3 L/min. if greater) and 100% E.O.S. and ±0.15 L/min for flow < 10% E.O.S (±0.1 L/min for flow E.O.S. ≤ than 2 L/min.)

Inlet pressures

3.5 - 5 bar ±20%

N2O blocking device

the supply of N2O is blocked when the O2 pressure drops below 0.4 bar ± 0.2 bar.

Charge loss level, side under pressure (before the adjusting needle valves)

less than 25 ml/min. in normal pressure conditions (ISO 5358)

Charge loss level, low pressure side (after the adjusting needle valves, vaporizer excluded)

less than 25 ml/min. at 30 cm H2O

Hypoxyguard system

at least 30% -3% O2 is garanteed in the mixture with N2O opening value equal to 0.4 L/min. (with inlet pressures of 3.5 bar)

Mixed gases outlet connection

conical connection 23 mm F ISO DIN 5356/1

Gases inlet connections

hose connector Ø 6 mm

O2 quick dosage
(if foreseen)

when pushed it can supply more than 45 L/min. O2

Download Area

Download here below the product sheet of this product. You can find any other information and document here.
Connections and Accessories

In compliance with the New Guidelines of the Italian Ministry of Health of 28/03/2013 regarding health advertising concerning medical devices, in vitro medical diagnostic devices and medical surgical facilities, we caution that the information contained in this website is exclusively directed at professional practitioners. All contents are of an informational nature and do not have advertising purposes.