RM flowmeters

The flowmeters type RM/185-2 are devices for the supply of medical gases with two meters suitable in respiratory therapy.

STRUCTURE
The measurement tubes are made in the body of the unit and the inlet connections are at the bottom of the units and arranged to be connected to the hospitals centralized gas plants or to gas cylinders completed with pressure regulators. For veterinary use, an anesthesia evaporator can be connected at the outlet of the dosage unit.

Tech Info

 

RM/185-2

Sizes (LxWxH)

93x86x205 mm

Weight

0.73 Kg

Dosage range O2 • N2O • Air

5 L/min. • 15 L/min. • 30 L/min. • 40 L/min.

Accuracy

±10% read value or ±0.5 L/min. if greater

Inlet pressures

3.5 - 5 bar ±20%

Charge loss level, side under pressure
(before the adjusting needle valves)

less than 25 ml/min. in normal pressure conditions (ISO 5358)

Charge loss level, low pressure side
(after the adjusting needle valves,
vaporizer excluded)

less than 25 ml/min. at 30 cm H2O

Mixed gases outlet connection

Ø 22 M. - 15 F. ISO DIN 5356-1 or conical connection 23 mm F ISO DIN 5356/1 (for veterinary use only)

Gases inlet connections

hose connector Ø 6 mm

Download Area

Download here below the product sheet of this product. You can find any other information and document here.
Connections and Accessories

In compliance with the New Guidelines of the Italian Ministry of Health of 28/03/2013 regarding health advertising concerning medical devices, in vitro medical diagnostic devices and medical surgical facilities, we caution that the information contained in this website is exclusively directed at professional practitioners. All contents are of an informational nature and do not have advertising purposes.