EN ISO 9170-2 AGSS
terminal units components
Painted stainless steel cover panel, for flush-mounted case for EN ISO 9170-2 AGSS outlet with front access for the outlet simplified calibration.
Flush-mounted case, complete with disposable panel, suitable for housing
EN ISO 9170-2 AGSS sockets and outlet.
EN ISO 9170-2 AGSS type 1L outlet.
VENTURI TYPE | Socket for EN ISO 9170-2 AGSS type 1 outlet with Venturi system. An injector supplies compressed air at 4 bars creating a constant suction (the setting value can be adjusted during installation) suitable to eliminate the anesthetic gases from the area where they are used. Front access for the simplified calibration without need to disassemble the outlet or to open the cover panel.
BLOWER TYPE | Socket for EN ISO 9170-2 AGSS type 1 outlet for blower.
A blower generates a constant suction inside the central system for the scavenging of anesthetic gases from the area where they are used. The flow can be adjusted during the installation directly on the socket by means of a variable section connector.
Probe for EN ISO 9170-2 AGSS type 1L outlet.
Pneumatic gauge indicating the terminal unit status, to be connected to the socket of the EN ISO 9170-2 AGSS outlet with Venturi system.
Surface mounted case, complete with varnished stainless steel cover panel, suitable for housing EN ISO 9170-2 AGSS sockets and outlets with front access for the outlet simplified calibration.
Download Area
Products Terminal Units
terminal units components