EASYMED® calibrated orifices oxygen flowmeters

EASYMED® flowmeters are top-quality instant flow measurement devices with calibrated orifices for regulating the dosage of oxygen and air in medical applications.

STRUCTURE
The flowmeters type EASYMED® are manufactured with an integrated pressure reducer for the supplied pressure stabilization and with the outlet fitting with thread adapted to the various required applications. The body is made of impact-resistant polymer with brass plated fittings, while a large ergonomic control knob allows an operator smooth drive for the selection of the gas supply value among the 10 possible options.

WORKING PRINCIPLE
The calibration of the flow is ensured by orifices got on a metal support with the laser technology. The small size structure and the particular technical configuration allow the EASYMED® “dial” flowmeters to be able to supply flows of medical gas with extreme accuracy even in the most difficult conditions, such as in emergency mobile units.

REGULATION VALUES FOR THE DIFFERENT END OF SCALE FLOWS
  STEP VALUES

END
OF
SCALES

6 L/min. 0,00 0,25 0,50 1,00 1,50 2,00 3,00 4,00 5,00 6,00
14 L/min. 0,00 0,50 1,00 2,00 4,00 6,00 8,00 10,00 12,00 14,00
15 L/min. 0,00 0,50 1,00 2,00 3,00 4,00 6,00 9,00 12,00 15,00
30 L/min. 0,00 1,00 2,00 4,00 8,00 12,00 16,00 20,00 25,00 30,00
50 L/min. 0,00 1,00 2,00 3,00 6,00 9,00 12,00 15,00 30,00 50,00
read more

Tech Info

Sizes (LxWxH)

82x54x65 mm

Weight

0.13 Kg

Supply pressure

280÷600 kPa with integrated pressure regulator for the supplied pressure stabilization

Accuracy

±10% read value or ±0.5 L/min. if greater between 10% and 100% E.O.S. 
±10% read value or ±0.2 L/min. if greater for flows < 10% E.O.S.

Standard supply connection

ISO G 1/4” M. • 1/4” NPT M.

Standard gas outlet connection

M12x1.25 M • 1/4” ISO 3253 M. • 3/8” ISO 3253 M. • 9/16” UNF EN 13544-2 M.

Gases options

O2 • Air

Standard full scale flow rate

6 L/min. • 14 L/min. • 15 L/min. • 30 L/min. • 50 L/min.

Flow rates

10 (0+9 preset values)

Flow calibration data

1013 mbar 23 °C

In compliance with the New Guidelines of the Italian Ministry of Health of 28/03/2013 regarding health advertising concerning medical devices, in vitro medical diagnostic devices and medical surgical facilities, we caution that the information contained in this website is exclusively directed at professional practitioners. All contents are of an informational nature and do not have advertising purposes.