EasyMED® PLUS calibrated orifices flowmeters

EasyMED® PLUS flowmeters are instant flow measurement devices with calibrated orifices for medical gases supply and dosage, particularly suitable in oxygen therapy. They are offered both in single or twin flowmeter execution to allow double and separate gas flow by using a single medical gas supply source.

STRUCTURE
The EasyMED® PLUS flowmeters are manufactured with an integrated pressure reducer for the supplied pressure stabilization and with an outlet nipple having two connection threads on the two ends for an easy and quick interchangeability. The operators can get two options of outlet thread by simply reversing the nipple, depending on the required applications. The body is made of impact-resistant polymer with brass chrome-plated fittings, while a large ergonomic control knob with Soft Grip inserts allows the operator to easily adjust the gas flow among the 10 possible preset options.

SIMPLE AND ACCURATE
The calibration of the gas flow is ensured by orifices got on a metal support with the laser technology. The I/0 switch button allows the operator to quickly lock and reactivate the flowmeter gas supply, keeping unchanged the previous flow preset value. The small size structure and the particular technical configuration allow the EasyMED® PLUS “dial” flowmeters to be able to supply flows of medical gas with extreme accuracy, even in the most difficult conditions, such as in emergency mobile units.

REGULATION VALUES FOR THE DIFFERENT END OF SCALE FLOWS
  STEP VALUES

END
OF
SCALES

6 L/min. 0,00 0,25 0,50 1,00 1,50 2,00 3,00 4,00 5,00

6,00

14 L/min. 0,00 0,50 1,00 2,00 4,00 6,00 8,00 10,00 12,00 14,00
15 L/min. 0,00 0,50 1,00 2,00 3,00 4,00 6,00 9,00 12,00 15,00
30 L/min. 0,00 1,00 2,00 4,00 8,00 12,00 16,00 20,00 25,00 30,00
50 L/min. 0,00 1,00 2,00 3,00 6,00 9,00 12,00 15,00 30,00 50,00
read more

Tech Info

Sizes (LxWxH)

61x104x78 mm

Weight

0.19 Kg

Gas supply pressure

280÷600 kPa with an integrated pressure reducer for the stabilization of supplied pressure

Gases options

O2 • Air

End of scale values

6 L/min. • 14 L/min. • 15 L/min. • 30 L/min. • 50 L/min.

Accuracy

±10% read value or ±0.5 L/min. if greater between 10% and 100% E.O.S.
±10% read value or ±0.2 L/min. if greater for flows < 10% E.O.S.

Standard inlet connection

ISO G 1/4” M. • 1/4”NPT M.

Standard outlet connection

Nipple with double thread, interchangeable by the end user:
• M12x1.25 M - 1/4” ISO 3253 M.
• 9/16” UNF EN 13544-2 M. - 1/4” ISO 3253 M.
• M12x1.25 M - 9/16” UNF EN 13544-2 M.

Flow calibration data

1013 mbar 23 °C

Flow rates

10 (0+9 preset values)

Watch the product video

EasyMED® PLUS calibrated orifices flowmeters

Discover the Easy Line

Excellence in Innovation, Breakthrough Technology, Intelligent Design

Download Area

Download here below the product sheet of this product. You can find any other information and document here.
Related products
Connections and Accessories

In compliance with the New Guidelines of the Italian Ministry of Health of 28/03/2013 regarding health advertising concerning medical devices, in vitro medical diagnostic devices and medical surgical facilities, we caution that the information contained in this website is exclusively directed at professional practitioners. All contents are of an informational nature and do not have advertising purposes.