EASYCARE® pressure regulators

The EASYCARE® two stage pressure regulator is suitable for direct mounting on oxygen cylinders.

STRUCTURE
Its structure is very compact and lightweight and it is particularly suitable for oxygen therapy. The body is made of brass and chrome-plated aluminum with an integrated impact resistant polymer calibrated orifices flowmeter with 10 options of flow oxygen supply. It is equipped with a pre-calibrated overpressure safety valve, and with a terminal unit (optional) connected to the first stage chamber which allows to get an additional and separate source of oxygen or air supplied at 4 bar. The inlet connection is offered in several options, as per the different reference standards of the destination Countries, and the outlet too is available with threads adapted to different applications.

WORKING PRINCIPLE AND GAUGE
The pressure calibration of the individual stages is fixed and a pressure gauge allows to get view of the cylinder contents. A protective silicone cover prevents gauge damages caused by possible shocks received during transport or use.

REGULATION VALUES FOR THE DIFFERENT END OF SCALE FLOWS
  STEP VALUES

END
OF
SCALES

6 L/min. 0,00 0,25 0,50 1,00 1,50 2,00 3,00 4,00 5,00 6,00
14 L/min. 0,00 0,50 1,00 2,00 4,00 6,00 8,00 10,00 12,00 14,00
15 L/min. 0,00 0,50 1,00 2,00 3,00 4,00 6,00 9,00 12,00 15,00
30 L/min. 0,00 1,00 2,00 4,00 8,00 12,00 16,00 20,00 25,00 30,00
50 L/min. 0,00 1,00 2,00 3,00 6,00 9,00 12,00 15,00 30,00 50,00
read more

Tech Info

Sizes (LxWxH)

147x55x117 mm (with UNI 11144 cylinder connection and without optional terminal unit)

Weight

0.81 Kg (without optional outlet)

Maximum inlet pressure

200 bar

Gases options

O2 • AIR

Standard flow range

6 L/min. • 14 L/min. • 15 L/min. • 30 L/min. • 50 L/min.

Flow rates

10 (0+9 preset values)

Flowmeter accuracy

±10% read value or ±0.5 L/min. if greater between 10% and 100% E.O.S. 
±10% read value or ±0.2 L/min. if greater for flows < 10% E.O.S.

Inlet connection to the cylinder

UNI 11144 • EN 850 • NF-E 29-650 
• BS 341-3 • DIN 477-1 • CGA

Outlet connection

1/4” ISO 3253 M. • 3/8” ISO 3253 M. 
• M12x1.25 M. • 9/16” UNF EN 13544-2 M.

Terminal Unit (optional)

AFNOR NF-E 29-650 • UNI 9507 • BS 5682 • DIN 13260-2 • SS 875 24 30

Download Area

Download here below the product sheet of this product. You can find any other information and document here.
Related products
Connections and Accessories

In compliance with the New Guidelines of the Italian Ministry of Health of 28/03/2013 regarding health advertising concerning medical devices, in vitro medical diagnostic devices and medical surgical facilities, we caution that the information contained in this website is exclusively directed at professional practitioners. All contents are of an informational nature and do not have advertising purposes.